What is the return policy?
For retail customers, there is a 90-day return policy. If you are not completely satisfied with your purchase, we will issue a full refund or exchange with a valid receipt. Customer damage or wear and tear is not covered unless a preexisting service contract states otherwise.
Read More
What is the relative intrinsic error of the devices?
RadTarge II D700, RadPavise, RadWall S300 Photon Beam Dose Equivalent Rate (µSv/h) Relative Intrinsic Error 137Cs – 662 keV 60.5 4.62% 121.0 0.93% 242.0 1.03% 484.0 1.00% 968.0 0.29% RadWall S100 Photon Beam Dose Equivalent Rate (µSv/h) Relative Intrinsic Error 137Cs – 662 keV 52 4.00% 499 -0.20% 1,490 -0.67%
Read More
What is the detection range of RadWall S?
At a distance of 4.9 ft (1.5 m), RadWall S can detect 1 mCi of Cs-137 within 2 s.
Read More
What is the energy response data?
RadTarge II D300 ISO Beam Tech Code Energy (keV) Response LK55 48 82.14% LK70 60 115.72% LK100 87 98.48% LK125 109 93.51% LK170 148 96.36% 137Cs 662 100.00% 60Co 1,250 64.86% RadWall S100 ISO Beam Tech Code Energy (keV) Response LK55 48 82.14% LK70 60 115.72% LK100 87 98.48% LK125 109 93.51% LK170 148 96.36% 137Cs 662 100.00% 60Co 1,250 75.00% RadTarge II D700, RadPavise, RadWall S300 ISO Beam Tech Code Energy (keV) Response LK35 30 82.00% LK55 48 89.01% LK70 60 112.74% LK100 87 107.03% LK125 109 83.78% LK170 148 90.12% 137Cs 662 100.00% 60Co 1,250 84.82% RadTarge II
Read More
How much data can each device hold?
RadTarge II, RadPavise Device: 43,800 data points (5 years at 1-hour intervals) Work log (5,000 items: turn on/off time, alarm on/off time) Software: Unlimited RadWall S Device: No need for data storage; real-time air absorption dose rates and alarming Software: Unlimited
Read More
Is there a manual for the calibration procedure?
Yes; the calibration manual will be available soon.
Read More
Is calibration software available?
Yes; calibration software is available to distributors and testing labs. RadTarge II, RadPavise RadTarge II D300, RadTarge II D700, and RadPavise are calibrated using calibration software on a PC via USB. RadWall S RadWall S100 and RadWall S300 are calibrated using calibration software on a PC via IoT-Cloud.
Read More
Does my device need calibration?
Devices do not need calibration to work properly. However, depending on the rules and regulations of your industry, you may be required to regularly calibrate your equipment for quality assurance purposes.
Read More
How can I tell if the battery is low?
RadTarge II, RadPavise The battery indicator on the screen shows the remaining charge on the battery. After running out of battery, the device continues to work for 4 hours. After 4 hours, the battery indicator flashes for 6 seconds before the device shuts down. RadWall S If there is no external power source, RadWall S will continue to run on its backup battery for 10 hours.
Read More