What is the relative intrinsic error of the devices?

RadTarge II D700, RadPavise, RadWall S300 Photon Beam Dose Equivalent Rate (µSv/h) Relative Intrinsic Error 137Cs – 662 keV 60.5 4.62% 121.0 0.93% 242.0 1.03% 484.0 1.00% 968.0 0.29% RadWall S100 Photon Beam Dose Equivalent Rate (µSv/h) Relative Intrinsic Error 137Cs – 662 keV 52 4.00% 499 -0.20% 1,490 -0.67%
Read More


FAQ Categories: Accuracy.

What is the energy response data?

RadTarge II D300 ISO Beam Tech Code Energy (keV) Response LK55 48 82.14% LK70 60 115.72% LK100 87 98.48% LK125 109 93.51% LK170 148 96.36% 137Cs 662 100.00% 60Co 1,250 64.86% RadWall S100 ISO Beam Tech Code Energy (keV) Response LK55 48 82.14% LK70 60 115.72% LK100 87 98.48% LK125 109 93.51% LK170 148 96.36% 137Cs 662 100.00% 60Co 1,250 75.00% RadTarge II D700, RadPavise, RadWall S300 ISO Beam Tech Code Energy (keV) Response LK35 30 82.00% LK55 48 89.01% LK70 60 112.74% LK100 87 107.03% LK125 109 83.78% LK170 148 90.12% 137Cs 662 100.00% 60Co 1,250 84.82% RadTarge II
Read More


FAQ Categories: Energy Response.