What is Hp(10) and Hp(0.07)?

Hp(10) and Hp(0.07) are personal dose equivalents, which represent the probability of stochastic health effects from radiation. Hp(10) is the deep dose equivalent. It is the equivalent dose at 10 mm below a specified point of the body. Hp(0.07) is the shallow dose equivalent, also referred to as skin dose equivalent or surface dose equivalent. It is the equivalent dose at 0.07 mm below the surface of the body. Currently, RadTarge II measures Hp(10). The option to switch between Hp(10) and Hp(0.07) will be added to future iterations of RadTarge II.
Read More


FAQ Categories: Software.

What is the energy response data?

RadTarge II D300 ISO Beam Tech Code Energy (keV) Response LK55 48 82.14% LK70 60 115.72% LK100 87 98.48% LK125 109 93.51% LK170 148 96.36% 137Cs 662 100.00% 60Co 1,250 64.86% RadWall S100 ISO Beam Tech Code Energy (keV) Response LK55 48 82.14% LK70 60 115.72% LK100 87 98.48% LK125 109 93.51% LK170 148 96.36% 137Cs 662 100.00% 60Co 1,250 75.00% RadTarge II D700, RadPavise, RadWall S300 ISO Beam Tech Code Energy (keV) Response LK35 30 82.00% LK55 48 89.01% LK70 60 112.74% LK100 87 107.03% LK125 109 83.78% LK170 148 90.12% 137Cs 662 100.00% 60Co 1,250 84.82% RadTarge II
Read More


FAQ Categories: Energy Response.