RadTarge

RadTarge is a series of all-digital electronic personal dosimeters that detect gamma and X-rays.

Features

  • YSO + SiPM and MVT
    Scintillation detector and patented DAQ algorithm
  • Direct reading
    Toggle between cumulative dose and dose rate
  • Real-time alarm
    Audible, visible, and vibrating alert options
  • Data logging and export
    Software available for PC
    » Download Software

Product Comparison

RadTarge-Mini RadTarge-H RadTarge-W
Recommended for strong radiation fields weak radiation fields strong radiation fields
Applications nuclear medicine (PET/CT centers), radiology departments, blood irradiation therapy, research labs and universities, baggage screening machine operators, pilots and flight attendants nuclear medicine (PET/CT centers), radiology departments, blood irradiation therapy, research labs and universities, baggage screening machine operators, pilots and flight attendants irradiation processing, industrial CT, industrial X-ray inspection, radioactive therapy, accelerator centers, nuclide production, nuclear power plants
Dose rate range 1 µrem/h–10 rem/h
(0.01 μSv/h–100 mSv/h)
1 µrem/h–1 rem/h
(0.01 μSv/h–10 mSv/h)
10 µrem/h–100 rem/h
(0.1 μSv/h–1 Sv/h)
Sensitivity 110 cps/mrem/h
(11 cps/μSv/h)
340 cps/mrem/h
(34 cps/μSv/h)
40 cps/mrem/h
(4 cps/μSv/h)
Alarm response time < 8 s < 2 s < 6 s
Energy range 30 keV–3 MeV 20 keV–3 MeV 20 keV–3 MeV
Energy Response ≤ ±30% @ 30 keV–1.5 MeV
≤ ±50% @ 1.5 MeV–3 MeV
≤ ±20% @ 20 keV–3 MeV ≤ ±30% @ 20 keV–1.5 MeV
≤ ±50% @ 1.5 MeV–3 MeV
Accuracy ±20% ±10% ±15%
Price $ $$ $$
RadTarge II D300 Details RadTarge II D700 Details RadTarge II D900 Details